Single Batch Calibration For Foam FM1-SB-05

FT1 SINGLE BATCH CALIBRATION SHEETS

The FT1 Dielectric Tensiometer measures water tension in soilless substrates. By default, it is supplied with standard foam discs, and a corresponding standard calibration curve is built into our DeltaLINK software.

 

An even higher accuracy option is available (when using the GP1 or GP2 Data logger) in the form of FT1-FM1-SB-05 foam discs. These “single batch” foam discs are supplied with a batch-specific colour coded calibration sheet, and It is very important that the correct calibration values are used.

Please NOTE:
When using the GP2 Data Logger, the supplied calibration sheet contains the required parameter values for the user to input into DeltaLINK Software. However, for the GP1 Data Logger, a different parameter value table is required. That table can be downloaded below.

HOW TO INSTALL THE SINGLE BATCH CALIBRATION FOR YOUR GP1 DATA LOGGER
  1. Close any open DeltaLINK windows.
  2. The downloadable lookup values are stored in a zip file. Download and open this.
  3. Extract the file with extension “.tbl”. Save it to your DeltaLINK Installation folder. Typically this might be “C:\Program Files (x86)\Delta-T\DeltaLINK 3.8” (note this will require administrator permissions on your PC)
  4. Open DeltaLINK, and configure the channel connected to the FT1 (1 or 2) as “custom voltage”.
  5. In the sensor’s “Input Channel Properties”, find the Linearization table drop-down menu, and select the lookup table.
  6. Fill in the items listed on the certificate. Change the units to kPa.
  7. Configure the temperature channel (3 or 4) as FT1 (temperature).

Download the GP1 table file here: FT1 Calibration Values (137 downloads )